LifeScan Recalls Specific Lots of Consumer and Professional OneTouch(R) SureStep(R) Test Strips Due To Inaccurate Readings at High Levels

MILPITAS, Calif., Feb. 24 /PRNewswire/ -- LifeScan, Inc. is conducting a voluntary recall in the United States of eight lots of OneTouch® SureStep® Test Strips, used by people with diabetes to measure their blood glucose levels at home.  The test strips are being recalled because they may provide falsely low glucose results when the glucose level is higher than 400 mg/dL.

The eight lots of consumer OneTouch SureStep Test Strips being recalled are:

Recalled Lot Size Description

# 2969251 100-ct OneTouch SureStep

# 2969798 100-ct OneTouch SureStep

# 2982369 100-ct OneTouch SureStep

# 2983467 100-ct OneTouch SureStep

# 2969795 50-ct OneTouch SureStep

# 2982566 50-ct OneTouch SureStep

# 2969481 50-ct Medicare/Mail Order

# 2998193 50-ct Medicare/Mail Order

Lot #'s are located on the outer carton and test strip vial.

Patients with test strips from the recalled lots are asked to call LifeScan at 800 574-6139 between 5:00 am and 7:00 pm Pacific Time, seven days a week or visit www.SureStep.com to request replacement product.  Replacement product will be shipped immediately and provided free of charge.

Advertisement

While waiting for replacement product to arrive, it is important that patients with recalled test strips continue to test their blood glucose.  Patients with access to a meter that does not use OneTouch SureStep Test Strips should use this other meter to test their blood glucose until replacement product from LifeScan arrives.  If an alternate meter is not available, patients may continue to test using the recalled OneTouch SureStep Test Strips.  However, if patients obtain results above 400 mg/dL, they should contact their healthcare professional for further instructions because their glucose may be significantly higher.

LifeScan estimates approximately fourteen thousand packages (50- and 100-count) of consumer OneTouch SureStep Test Strips were distributed nationwide between August 1, 2009 and January 28, 2010.

No injuries associated with these recalled test strips have been reported.  However, if patients use the falsely low test results to determine their insulin dose, they may give themselves too little insulin, which could result in poor blood glucose control.  Patients with high blood glucose may or may not have certain symptoms including increased thirst, frequent urination, headaches, difficulty with concentration, blurred vision and fatigue (weak, tired feeling).  High blood glucose must be recognized and treated promptly to avoid serious complications, such as coma and death.

Hospitals, clinics and other multi-patient facilities using SureStep®Pro®, SureStep®Flexx® or OneTouch® SureStep® Hospital Systems have also been notified of this issue.  All three of these systems use SureStep®Pro® Test Strips which also may provide inaccurately low test results when the blood glucose reading is greater than 400 mg/dL.  LifeScan has advised these facilities of the appropriate actions to take in order to continue to use these SureStepPro Test Strips.  However, healthcare facilities with access to alternative means of testing should consider temporarily discontinuing their use of their SureStep Systems until replacement test strips are available.

This field action is limited to eight lots of OneTouch SureStep Test Strips sold for consumer use in the U.S., and select SureStepPro Test Strips used in healthcare facilities. Similar recall actions are being taken in other countries where the affected product was distributed.

In 2006 LifeScan stopped selling OneTouch SureStep Meters in the U.S. but continued to provide test strips.  Today, OneTouch SureStep Meter users represent a very small portion of LifeScan's total customer base.  The vast majority of LifeScan's customers now use OneTouch® Ultra® Brand Meters, which use an entirely different technology.

LifeScan, Inc. is a leading maker of blood glucose monitoring systems for people with diabetes. For information about diabetes care and LifeScan products and services, visit www.OneTouchDiabetes.com.

 

 

Advertisement


 

Bookmark and Share | Print | Category | Home

This article originally posted 25 February, 2010 and appeared in  Issue 510Safety and Error PreventionBG ControlMedical Devices

Past five issues: Diabetes Clinical Mastery Series Issue 252 | SGLT-2 Inhibitors Special Edition July 2015 | Issue 792 | Diabetes Clinical Mastery Series Issue 251 | Issue 791 |

2015 Most Popular Articles:

Self-Monitoring of Blood Glucose: The Patient's Perspective
Posted July 10, 2015
The Impact of Glucagon-Like Peptide 1 Receptor Agonists on Weight Reduction
Posted July 10, 2015
Vitamin D3 Supplementation and Weight Loss Prevent Cancer in Older Women
Posted July 10, 2015
GLP-1 Receptor Agonists Reduce Bone Loss During Weight Loss
Posted July 17, 2015
Diet That Mimics Fasting Appears to Slow Aging
Posted July 03, 2015
Metformin May Increase Mortality in Advanced CKD among Type 2 Patients
Posted July 10, 2015
Evaluation of Cellphone Application that Records Readings from Glucometer
Posted July 10, 2015
Impact of Exercise in Type 2 Diabetes Patients
Posted July 23, 2015
Ancestral Diets Can Determine Vulnerability to Type 2 Diabetes
Posted July 17, 2015
Two Gene Mutations Identified in Increased Risk of Obesity and Diabetes
Posted July 23, 2015


Browse by Feature Writer & Article Category.
A. Lee Dellon, MD | Aaron I. Vinik, MD, PhD, FCP, MACP | Beverly Price | Charles W Martin, DD | Derek Lowe, PhD | Dr. Brian Jakes, Jr. | Dr. Fred Pescatore | Dr. Tom Burke, Ph.D | Eric S. Freedland | Evan D. Rosen | Ginger Kanzer-Lewis | Greg Milliger | Kristina Sandstedt | Laura Plunkett | Leonard Lipson, M.A. | Louis H. Philipson | Maria Emanuel Ryan, DDS, PhD | Marilyn Porter, RD, CDE | Melissa Diane Smith | Michael R. Cohen, RPh, MS, ScD, FASHP | Paul Chous, M.A., OD | Philip A. Wood PhD | R. Keith Campbell, Professor, B.Pharm, MBA, CDE | Richard K. Bernstein, MD | Sheri R. Colberg PhD | Sherri Shafer | Stanley Schwartz, MD, FACP, FACE | Steve Pohlit | Steven V. Edelman, M.D. | Timothy S. Hollingshead |

Cast Your Vote
Do you have any patients on a dual therapy of a DPP-4 and a SGLT-2 who have reached their goal A1c?

CME/CE of the Week
Calorie Expenditure: Raising Your Clients' Awareness
Anne Daly, MS, RD, BC-ADM, CDE  
Category: Nutrition
CE Credits: 50